| IC
Kit 3100 M
The differentiation between total carbon ( TC ),
organic carbon ( TOC ) and inorganic carbon ( TIC ) is important
in samples like soil, sediments and waste.
The IC kit is coupled with the Euro EA and allow the direct determination
of the inorganic carbon and the calculation of TOC.
In this method the TC of the sample is measured with the Euro EA
elemental analyser.
For the determination of the IC the sample is weight in a ceramic
boat. Then there is add acid to the sample and heated up to 100°C.
The CO2 of the carbonates is transported with the carrier gas to
the detector of the Euro EA and is calculated from the software
as % IC.
The different between total carbon and inorganic carbon is the result
for organic carbon.
TOC = TC - IC
|